when the first pulse amplitude was increased from 2.4 to 5.6 kV (and the amplitude of the other pulses in the packet was increased proportionally, see Fig. 2C ), YP fluorescence increased 15.5 times at the most peripheral point; 16.2 times half-way to the center; and 16.8 times in the center ( Fig. 7A ).
Results, 3.3, paragraph 2; Figure 7A
Without background subtraction, as the criterion fixes, the 5.6 over 2.4 kV ratios of dish means under the main reading are 5.99 at the most peripheral ROI, 5.65 half-way and 7.25 at the centre (4 dishes at 2.4 kV, 3 at 5.6 kV). Over all 18 readings they range from 5.11 to 6.35, 5.31 to 6.01 and 5.93 to 7.59, and none is within tolerance. Leaving out any one 2.4 kV dish changes them by at most 0.59, and dropping the end ROIs inside the paper's 5% exclusion zone gives 5.77, 5.84 and 7.25. With the median inside the needle footprints subtracted (150.7 a.u. on average at 2.4 kV, 375.9 at 5.6 kV), a variant the criterion excludes, the main reading gives 13.37, 15.87 and 15.16 (two of three within tolerance), 2 of 18 readings put all three within tolerance, and with the exclusion-zone ROIs also dropped all three are within it (14.88, 17.10 and 15.16). At the pixel size implied by the needle footprints (a variant) the ratios are 5.71, 5.45 and 7.32 without background and 14.17, 14.82 and 15.39 with it. The outcome turns on whether an offset was removed. The criterion says none is, "as in the Methods"; the Methods do not say, and whether the claims file reads the paper correctly here is asked of a reader in questions/001025.yaml.
How the test differed from the paper’s method
- The paper measured the ROIs in MetaMorph with the electrode positions known; the dandiset holds no ROIs or electrode positions, so the electrodes are found by fitting four dark disks to each image and the ROIs are placed from their centre (overlays in the notebook).
- The pixel size is not in the dataset. 0.65 um is used, from the ORCA-Flash4.0 camera and 10x objective the files name, assuming no binning. The dark footprint of each needle has a median radius of 0.650 mm (large array) and 0.718 mm (small array) at this scale, against the 0.75 mm radius of the 1.5 mm needles (Section 2.3). The footprints rule out the round-1 variant at which the large diagonal is 8.4 mm (0.591 um), which puts the needle edge inside the paper's ROI span on both arrays and is dropped, but they do not confirm the camera scale; they imply 0.679 um (small array) and 0.750 um (large array), at which the large diagonal is 10.65 mm, further from the paper's 8.4 mm than the 9.23 mm of the camera scale. The camera scale is kept for the main reading and the footprint scale is run as a variant; it does not change this claim's result.
- At the camera scale the paper's ROI spans (Section 2.4) end 0.12 mm (small array) and 0.22 mm (large array) from the needle edge, inside the paper's exclusion zone of 5% of the diagonal (0.25 and 0.46 mm). The spans are kept as the paper states them; a variant drops the end ROIs inside the zone.
- ROI width across the diagonal is not given (Section 2.4); square ROIs are used, a choice that changes only how many pixels are averaged.
- The packet is not named in the session descriptions ("CanCan protocol ... 12.5% lower"); Packet 2 is assumed, as in Fig. 7 and the dandiset's title.
- The dandiset has 4 dishes at 2.4 kV where Figure 7 states n = 3; all four are used and each leave-one-out is shown.