In control cell monolayers exposed to a train of five single pulses applied to one of the base electrodes (833 kHz, 4 kV), the distribution of YP fluorescence ( Figure 4 A) simply followed the electric field strength map
Results, Section 2.3, paragraph 1; Figure 4A
Over 2627 ROIs of 200 um within 6 mm of the image centre, with E1 taken as the lower-right base electrode (base_a), as Figure 4A labels it, background-subtracted YP and the computed field have Spearman rho = 0.666 (p printed as 0 for the full grid, which treats contiguous ROIs as independent). On ROIs thinned to 1 mm spacing rho = 0.666 (p = 9.3e-15, 105 ROIs), and to 2 mm rho = 0.654 (p = 0.00053, 24 ROIs). The order of the three electrode neighbourhoods (0 to 500 um from the footprint edge) is the same by YP (base_a 649.8, base_b 262.6, apex 82.2) and by field (12.88, 7.38, 5.31 kV/cm). Across grid radii of 5 and 6 mm, margins of 150 and 400 um and neighbourhoods of 250, 500 and 1000 um, rho is 0.640 to 0.746 and the order always matches; p at 1 mm spacing is 2.7e-8 or less, and at 2 mm spacing below 0.05 in 8 of 10 variants (0.095 in the two with a 5 mm grid and a 400 um margin, 12 ROIs). With E1 taken as the other base electrode, rho is 0.424 to 0.510 (0.397 to 0.643 at 2 mm spacing, p up to 0.054) and the order never matches, so the criterion's two readings would disagree. Where Figure 4A places E1 is not shown in the notebook, so the assignment rests on the uptake (base_a has the most); the outcome also rests on this file being the image of Figure 4A, which its session description contradicts (findings). The question of which electrode was E1 is put to the paper's authors (findings). Outside the entry, the author of this revision mapped the deposited images onto the published Figure 4 image (PMC10341945) through the three electrode centres; the registry's checks keep a published figure out of the notebook, so this is not shown there and is not verified. The paired-pulse file matched Figure 4B in fine detail, turned by 180 degrees and not mirrored (correlation of fine detail 0.18, 0.34 after a one-pixel shift); the single-pulse file did not match Figure 4A in either orientation (at most 0.03 over shifts, rotations and scalings of the map, the level of mismatched pairs). If that holds, the image of Figure 4A is not deposited and the claim would be blocked (data missing); no change is proposed, as the analysis cannot be in the notebook.
How the test differed from the paper’s method
- The field map is computed here by a two-dimensional finite-difference solve (40 um grid, electrodes as 1.5 mm discs of fixed potential, insulating wall of a 35 mm dish centred on the image centre), in place of the paper's Sim4Life model (Methods 4.3, "1.6 S/m solution to a depth of 3.5 mm", field "in the plane 1 um above the glass"). The needles span the solution from the glass up (Methods 4.2, "lowered ... until the tips touched the glass bottom"), so the field does not vary with height. Electrode positions are measured in the image, and the scale is set from the 1.5 mm needle diameter (Methods 4.2); it gives 0.654 um per pixel, the camera's 0.65.
- Assumed that E1, the electrode pulsed in Figure 4A, is the lower-right base electrode of the file's image (base_a). The file does not record it, and the paper's text does not place it (the Figure 4 legend says only that the panels are shown "with electrode positions labeled"). The published panel A has the apex at the top and E1 at the lower right, as the deposited image has the apex at the top, but that is not shown in the notebook, and the deposited images do not keep one orientation (the paired-pulse image has its apex at the bottom). base_a is also the base electrode with the most uptake (with E1 = base_b the grounded electrode would show 2.5 times the uptake of the pulsed one), so the assignment rests on the uptake. The notebook shows both assignments.
- The single-pulse file is taken as the image of Figure 4A from its name, the dandiset's title (the Figure 4 legend) and its layout (three electrodes in a triangle, uptake around one base electrode). Its session description gives another exposure (four co-directional paired-pulse trains at 770 kHz, 10 kV/cm, "the electric field direction stayed the same in each pair"). It does not say which electrodes were energized; that the field came from one base electrode is an inference from the uptake around one of them.
- The ROI grid (within 6 mm of the image centre, leaving out ROIs within 150 um of a footprint edge) and the neighbourhood (0 to 500 um from the footprint edge) are choices the claims file does not fix; the notebook shows 5 and 6 mm, 150 and 400 um, and 250, 500 and 1000 um, and the outcome is the same under all of them. The Spearman test is also run on ROIs thinned to 1 and 2 mm spacing, as contiguous ROIs are spatially dependent.
- The amplitude is taken as 4 kV (Section 2.3); the file's description says 2 kV. A rank correlation and an order do not depend on the amplitude.